Operating instructions
University of Saskatchewan 
Electrical Engineering Laboratory Equipment Manual 
The display of the analyzer may be set to start or center modes, as indicated by 
the  switch  with  STR  and  CTR  above  it.    Depending  on  this  switch,  the  LED  display 
indicates  either  the  start  frequency  in  Hz,  corresponding  to  the  leftmost  line  on  the 
display, or the center frequency in Hz, corresponding to the center line on the display. 
The  adjust knobs  (coarse/fine) to the  right of  the LED display control the  start/center 
frequency.  Not all of the HP 3580As have this LED  readout;  some have an analog 
start/center  frequency  readout.    It  doesn’t  matter  what  type  you  choose,  as they both 
work equally well. 
The SWEEP MODE dial should be set to REP (repetitive). At no point in the labs 
will  you  need  to  set  the  mode  to  anything but  repetitive.   The  remaining  three dials, 
RESOLUTION BANDWIDTH, FREQ SPAN/DIV, and SWEEP  TIME/DIV all work in 
conjunction.   As  was  mentioned  earlier,  selecting  a  narrow  RBW  will  yield  a  more 
accurate spectrum, but the sweep speed will need to be slowed in order to maintain an 
accurate  reading.    If  your  selected  sweep  speed  is  too  fast  to  maintain  an  accurate 
reading, the ADJUST indicator will light.  Adjust the sweep speed to a slower value until 
the ADJUST indicator turns off. At that point, the display will be accurate. 
The  RESOLUTION  BANDWIDTH  dial  has  another  control,  DISPLAY 
SMOOTHING built into its knob.  This control has three settings, and should be rotated 
fully counterclockwise to its minimum position.  Display smoothing is also known as the 
video  bandwidth  on  modern  analyzers,  and  is,  as  the name  suggests,  a  control  that 
adjusts  the  smoothness  of  the  displayed  spectrum.    You  may  think  that  the  display 
smoothing should be set to its maximum setting all the time, but like the other frequency 
sweep controls, it also is tied into the sweep speed.  At its maximum setting, the sweep 
speed must be set prohibitively slow to ensure an accurate display. 
The amplitude controls are found along the bottom of the instrument. The 
analyzer  has  two  vertical  display  modes:    linear  and  logarithmic.    These  modes  are 
selected  by  the 
AMPLITUDE MODE 
pushbuttons  found 
just  to  the  right  of 
the power switch. 
The UNCAL indicator light found to the right of the INPUT SENSITIVITY knob will 
light  when  the  CAL  adjust  knob  is  not  rotated  fully  clockwise.    This  means  that  the 
display of the analyzer is uncalibrated, and therefore inaccurate. Always rotate the CAL 
knob fully clockwise so that the UNCAL indicator stays off. 
Before ever connecting an input signal to the analyzer, always adjust the INPUT 
SENSITIVITY  control  to  its  least  sensitive  setting.    This  means  the  fully 
counterclockwise position, minus one.  The fully counterclockwise position is the CAL 
(calibrate) mode.  This mode disconnects the front panel inputs from the analyzer, and 
internally connects a 10 kHz calibration signal, obviously for calibration purposes.  This 
mode will be explained in detail later. 
LIN
10 dB
1 dB
LINEAR
LOG
dB / DIV
ON (AC)
OFF
ON (BAT)
 CHARGE
dBv / LIN
dBm
600
CAL
10 kHz
NORMAL
CAL
MAX
INPUT
OVERLOAD
UNCAL
VERNIER
(CASE GROUND)
POWER AMPLITUDE MODE AMPLITUDE REF LEVEL INPUT SENSITIVITY INPUT










