Operating instructions
University of Saskatchewan 
Electrical Engineering Laboratory Equipment Manual 
Printing the Display 
The HP 3580A is an analog spectrum analyzer and cannot directly print to the 
printers  available  in  the  laboratories.    However,  you  can  hook  the  analyzer  up to  an 
Oscilloscope, and print the spectra from there. 
On  the  back  of  the  analyzer  are  a  number  of  connectors.    The  ones  marked 
RECORDER X-AXIS and Y-AXIS are the ones you will need.  These outputs are meant 
for printing to an old-fashioned pen plotter. 
The oscilloscope should have its two X and Y input channels (channels 1 and 2 
respectively) set to dc coupling, and 500mV/div.  Position each channel’s ground level 
to approximately the middle of the screen. The X-AXIS output on the analyzer goes to 
the X channel on the scope, and similarly the Y-AXIS output gets connected to the Y 
channel. Once you have made the necessary connections, turn on the X vs. Y display 
mode of the oscilloscope. 
You  should  see  a  dot  moving  across  the  Scope  screen.    Set  the  Analyzer’s 
Sweep Mode knob to MAN for manual. Use the Sweep Mode vernier knob to adjust the 
dot on the analyzer’s display to the center vertical graticule.  Use the Scope X channel 
position knob to move the dot left or right, and use the Scope Y channel position knob to 
move it up or down.  Your goal is to place the dot on the center vertical graticule of the 
scope display to resemble the dot on the analyzer’s display.  Once this is done, set the 
Sweep Mode knob to SING for single sweep. 
Once this has been accomplished, press the Auto-store button (HP Scopes) or 
set the display persistence to infinite (some of the other scopes). Press and release the 
Display Clear Write button, on the Analyzer, to activate a single display sweep. Once 
the spectrum analyzer’s screen has been traced by the scope, press the Scope’s stop 
button and print the waveform. 
Amplitude Response of Two-Port Networks 
There is an output on the back of the spectrum analyzer called the TRACKING 
OSC (oscillator) OUT, along with a LEVEL adjust knob.  The output is a sine wave with 
a variable swept frequency that is synchronized to, or tracks the frequency being 
analyzed across the screen at any point in time. This is used to make a measurement of 
the frequency response of a two-port network (i.e. a filter, an amplifier, etc.). The output 
impedance of the TRACKING OSC OUT is 600  and not the standard 50  found in all 
signal  generators.    For  this  reason,  a  buffer  amplifier  (a  voltage  follower)  should  be 
employed  to  isolate  the  tracking  oscillator  from  the  network  you  are  testing  to  avoid 
circuit  loading.  Connect  the  TRACKING  OSC  OUT  through  a  voltage  follower  to  the 
input  of  the  spectrum  analyzer.  The  TRACKING  OSC  LEVEL  knob  is  adjusted for  a 
reference level of aprox. - 40 dBv on the front display. All measurements are referenced 
from this level. Connect the output of the voltage follower to the input of your network. 










