the time it takes the FFT processor to calculate the spectrum. With narrow spans, where 
the time record is long then the processing time, the processor would calculate one FFT 
per time record and wait till the next time record  is completed. The update would be 
slow. With overlap processing, data from the previous time record and the current time 
record is used to calculate the next FFT. The overlap is a percentage of the time record. 
No  overlap  is  0%  and  99.8%  reuses  511  out  of  512  samples.  The  SR700  uses  the 
maximum  amount  of  overlap  allowed.  This  varies  with  the  span.  The  analyzer  uses 
overlap processing in the continuous trigger mode only. No overlap is allowed, on spans 
of 25KHz and over. The average menu may be used to enter a smaller overlap. 
( Standford Research Systems, 2000 ) 