Operating instructions
University of Saskatchewan 
Electrical Engineering Laboratory Equipment Manual 
large number of averages, in the exponential mode, will not capture any changes in the 
signal. 
The  SOURCE  menu  selects  the  output  waveform.  The  Configure  Source 
submenu sets the parameters. The 2-Tone waveform is the sum of two pure sine waves 
for intermodulation distortion tests. 
The  Chirp  source  provides  an  equal  amplitude  sine  wave  at  each  bin  of  the 
span.  The  phases  of the  sine  waves are  set  so  that  they  do not  add  in  phase. This 
source  is  used to measure  transfer  functions  (frequency  response).  The waveform  is 
periodic over the time record and does not require windowing. Use the Uniform window 
and the Src (source) trigger. Turn the Auto Offset off in the input menu. The source is 
digitally  synthesized  and  has  an  output  filter.  The  ripple  of  this  filter  appears  at  the 
output.  Source  Cal  compensates  for  the  output  ripple  by  adjusting  the  input 
calibrations.  The  signal  at  the  source  output  is  not  adjusted.  If  a  device  is  inserted 
between the source and the input, the amplitude frequency response of the device will 
be displayed. The Auto Phase key performs the auto phase function. With Src (source) 
trigger, the phases of the sine waves of the source are stable but random. A calibrated 
source phase curve is required to measure the phase of a device under test. The auto 
phase function measure the current phase spectrum of the source and then subtractes 
it  from  the  next  phase  spectrum  to  remove  the  phase  from  the  source.  With Auto 
Phase, the phase spectrum will be 0 degrees across the span. If a device is inserted 
between the source and the input, the phase spectrum will be the phase response curve 
of the device under test. Auto Phase is turned off whenever the span or source type is 
changed. 
NOTE:    A inherent  problem  exists  when  a frequency  response  is  done  with  a 
small linewidth. Errors may occur if the start frequency is not at zero or Center (on the 
marker  keypad) is  used.  Detail  expansion of  the frequency  span can  be  done  using 
Expand X (zoom) with no errors. Ensure the frequency of interest is at the center of a 
bin. It is always best to know the expected output. An analog spectrum analyzer with a 
sine source is a more accurate analyzer to use for a frequency response. 
Frequency Response Summary 
-  Use Chirp source. 
-  Uniform window. 
-  Source Cal. 
-  Trigger Src. 
-  Auto Offset off ( Input menu). 
-  Auto Phase optional. 
-  Start frequency at zero. 
-  Do not use Centre on Marker keypad. 
-  Place Marker on centre frequency and use Expand X.    










