Operating instructions
University of Saskatchewan 
Electrical Engineering Laboratory Equipment Manual 
  Vertical Current/Div: 2 mA/div 
Termination Knob (on the right side of the shelf): 
  Base Term: Step Gen (fully clockwise position) 
As with the 575s, before you actually switch in the transistor to the instrument, 
you  must  make  sure  that  the  baseline  is  aligned  properly.    Using  the  horizontal  and 
vertical  position  controls,  align  the  leftmost  point  of  the  baseline  so  that  it  is  directly 
underneath the bottom leftmost crosshair on the display. 
To test a diode, the same settings as for an npn transistor are used, except the 
horizontal volts/div should be reduced to 0.2 volts/div.  Connect the diode to the C and 
E terminals of the test pod, just as with the 575s. 
To test a pnp transistor, the settings outlined above for an npn transistor are also 
used, with the following two exceptions: 
  Collector Supply Polarity: - (PNP side) 
  Display: NORM pushbutton in out position (invert) 
With the display inverted, the baseline is aligned to the bottom left side of the screen, as 
with npn transistors.  However, now the vertical axis is –I
C
, and the horizontal axis is –V
CE
. The 
origin is still located at the bottom left corner of the display. 