5) Three controls need to be adjusted to give a satisfactory display: the Vertical Section
volts/div and position, and the Horizontal Section time/div. These controls can be
adjusted two ways: by YOU, or by the oscilloscope.
For now, let’s have the oscilloscope do the adjustment. Press the Autoscale hardkey, a
white button just above the Vertical Section. You should see a display very similar to the
one below with about 2_ cycles of the squarewave (although the corners may not look quite
so square as they do below).
Note that a softkey called Undo Autoscale has appeared at the lower left of the display.
This can be a very valuable and time-saving button to use, under certain circumstances. Try
“undoing autoscale” by pressing this softkey button. Then, press the Autoscale hardkey
again, and you should see about 2_ cycles of the squarewave.
The next step (number 6, Probe Compensation) is so important it has an entire
page devoted to it.