Specifications
To trigger on a specific line of video
TV triggering requires greater than 1/4 division of sync amplitude, either
channel 1 or channel 2 as the trigger source. Turning the trigger level knob
in TV trigger does not change the trigger level because the trigger level is
automatically set to the sync pulse tips.
One example of triggering on a specific line of video is looking at the vertical
interval test signals (VITS), which are typically in line 18. Another example
is closed captioning, which is typically in line 21.
1
Select the TV display, TV as the trigger mode, and the appropriate
TV standard.
2
Press 
 Slope/Glitch 
 in the 
TRIGGER 
section of the front panel,
then press the 
TV
 and 
TV Menu
 softkeys. Press 
Line
. Select the
number of the line you want to examine.
3
Press the 
Trigger On Field
 softkey to select the TV field of the line you
want to trigger on. Your choices are 
Field 1
, 
Field 2
, and 
Alt Fld
(Alternate Fields). 
Alternate Triggering
If Alt Fld is selected, the oscilloscope will alternately trigger on the selected line
number in Field 1 and Field 2. This is a quick way to compare the Field 1 VITS
and Field 2 VITS or to check for the correct insertion of the half line at the end of
Field 1.
When using 
GENERIC
 as the TV standard, the Trigger On Field softkey gives
you the choices of Field 1, Field 2 and Vertical.
Using Option 005 Enhanced TV/Video Trigger (HP 54645A)
To trigger on a specific line of video