Specifications
To check the output of the CALIBRATOR
In this test, you measure the output of the 
rear-panel DC CALIBRATOR
OUTPUT with a multimeter. The
 CALIBRATOR
 is used for self-calibration of
the HP 54645A/D. The accuracy of the 
CALIBRATOR
 is not specified, but it
must be within the test limits to provide for accurate self-calibration.
The 
CALIBRATOR
 output produces DC voltages between 0 and 5V during
vertical calibration.
Test limits: 
  DC calibrator:  5.000 V 
±
10 mV and 0.000 V 
±
 500 
µ
V
Table 7-4
Equipment Required
Equipment Critical specifications Recommended
Model/Part
Digital Multimeter 0.1 mV resolution, better than 0.01% accuracy HP 34401A
Cable BNC HP 10503A
1
Connect a multimeter to the rear-panel 
CALIBRATOR
 connector.
2
Press 
 Print/Utility 
.
3
Press the 
Service Menu
 softkey, then the 
Self Test 
softkey, and then the
DAC
 softkey.
The multimeter should measure 0.00 V dc 
±
 500 
µ
V. If the result is not
within the test limits, see "Troubleshooting the Oscilloscope."
4
Press any key once to advance the test.
The multimeter should read 5.000 V 
±
 10 mV. If the result is not within the
test limits, see "Troubleshooting the Oscilloscope."
5
Press any key once to end this test.
Testing, Adjusting, and Troubleshooting
To check the output of the CALIBRATOR
7-15










