Specifications
1
Check the rear panel 
CALIBRATOR
 output level.
If you are not sure how to check the 
CALIBRATOR
, see "To check the output
of the 
CALIBRATOR."
2
Load the default calibration factors.
a
Set the rear-panel 
CALIBRATION
 switch to the 
UNPROTECTED
position.
b
Press
 Print/Utility 
. Then press 
Service Menu
 and
Self Cal Menu
.
c
Press the 
Load Defaults
 softkey.
Vertical 
self cal
3
After the message "Default scope calibration factors have been
loaded" is displayed on the lower side of the display, press 
Vertical
.
4
Follow the instructions on the display, then press 
Continue
.
5
After the message "Press Continue to return to calibration menu"
appears on the display, press the 
Continue
 softkey
Delay self cal 
for HP 54645A 
Only
6
Connect a pulse generator set to 100 kHz and 1 V
p-p
, with a rise time
of less than 10 ns, to channels 1 and 2. Place the feedthrough at the
BNC tee. Make sure you use the HP 10502A cables to ensure equal
cable lengths.
7
Press the 
Delay
 softkey, then follow the instructions on the display.
8
Set the rear-panel 
CALIBRATION
 switch to 
PROTECTED
.
Delay self cal 
for HP 54645D
Only
9
Connect channels A1 and A2 to the front-panel probe adjust signal
(5 V
p-p
, 1.2 kHz) using two 10:1 probes (HP 10074A). Connect the
ground leads on the 10:1 probes to the front-panel ground lug using
the banana ground plug (HP 1252-7310).
10
Connect digital channel D0 to the front-panel probe adjust signal
(5 V
p-p
, 1.2 kHz) using the 16-channel cable (HP 54620-61801).
Connect the ground lead of this cable to the front-panel ground lug.
11
Press the Delay softkey, then follow the instructions on the display.
12
Set the rear-panel 
CALIBRATION
 switch to 
PROTECTED
.
Testing, Adjusting, and Troubleshooting
To perform the self-calibration
7-34