Specifications
To operate the time base controls
The following exercise guides you through the time base keys, knobs, and
status line.
•
Turn the Time/Div knob and notice the change it makes to the status
line.
The Time/Div knob changes the sweep speed from 2 ns to 50 s (HP 54645A)
and from 5 ns to 50 s (HP 54645D) in a 1-2-5 step sequence, and the value is
displayed in the status line.
•
You can access the horizontal Vernier softkey by pressing the 
 Main/Delayed 
 key.
The Vernier softkey allows you to change the sweep speed in smaller
increments with the Time/Div knob. These smaller increments are
calibrated, which result in accurate measurements, even with the
vernier turned on.
•
You can check the sample rate by pressing the 
 Main/Delayed 
 key
when the instrument is running, during a single run, or when the
instrument is stopped. The sample rate is also displayed when the
acquisition stops or ends.
•
Turn the Delay knob and notice that its value is displayed in the status line.
The Delay knob moves the main sweep horizontally, and it pauses at
0.00 s, mimicking a mechanical detent. At the top and bottom of the
graticule is a solid triangle ( 
▼
 ) symbol and an open triangle ( 
∇
 )
symbol. The 
▼
 symbol indicates the trigger point and it moves with
the Delay knob. The 
∇
 symbol indicates the time reference point. If
the time reference softkey is set to left, the 
∇
 is located one graticule
in from the left side of the display. If the time reference softkey is set
to center, the 
∇
 is located at the center of the display. The delay
number tells you how far the reference point 
∇
 is located from the
trigger point 
▼
.
Front-Panel Overview
To operate the time base controls
2-27