User`s guide

6 Sample Sessions: Other Mini-OTDR Modes How to set up a Multi Fiber Test
156 E6000C Mini-OTDR Users Guide, E0302
9 Add your own file extension by entering a . followed by
characters of your choice.
If you do not specify a file extension, the default
extension .SOR is used.
NOTE File names are a maximum of 8, and file extensions are a
maximum of 3 characters long.
File name limits If you attempt to add to an 8 character file name or to a 3
character extension, you will hear a beep and no
characters will be added.
How to save multiple settings
10 If you want to measure the fiber with more than one
setup, repeat step 2 (and steps 6 to 9 if required) for
the remaining Measurements (and Save As) edit
fields.
NOTE If you choose the same name for more than one setting,
the first measurement is saved to a file, and this file is
overwritten when the second measurement is taken.
So, for example, if you save two separate measurements
as USER_01.SOR, the first measurement will be saved to
the file USER_01.SOR, and this file will be overwritten
when the second measurement is taken.
How to Set the Multi Fiber Test Trace
Information
Trace Info screen 1 Navigate to the Multi Fiber Test Trace Info screen.
See “How to Navigate within the Multi Fiber Test
Config page” on page 152 for details of how to navigate.
You see the Trace Info screen (Figure 60).