User`s guide

How to Use the Visual Fault Finder submodule 6 Sample Sessions: Other Mini-OTDR Modes
E6000C Mini-OTDR Users Guide, E0302
171
How to Take the Measurement
Insert DUT 10 Insert the Device Under Test in the link between the
Source and the Power Meter (Figure 70, second
picture).
NOTE Figure 88 on page 260 shows how you might set up the
Device Under Test (DUT) for measurements using the
power meter.
11 Switch on the Source.
Select On/Off from the Source window.
Read Insertion Loss 12 Read the insertion loss for the DUT from the Power
Meter window.
13 Switch off the Source.
Select On/Off from the Source window.
14 Disconnect the DUT.
How to Use the Visual Fault
Finder submodule
Insert submodule 1 Switch off the Mini-OTDR, and insert a module. Insert a
6007A Visual Fault Finder submodule into the
submodule slot in the module (see “Inserting and
Removing a Submodule” on page 74).
Connect fiber 2 Attach the required optical connector interface to the
optical output.
3 Connect the fiber to this interface.