A Installation and Maintenance AC Line Power Supply Requirements
178 E6000C Mini-OTDR Users Guide, E0302
WARNING To avoid the possibility of injury or death, you must
observe the following precautions before switching on the
instrument.
If this instrument is to be energized via an
autotransformer for voltage reduction, ensure that the
common terminal connects to the earth pole of the
power source.
Insert the power cable plug only into a socket outlet
provided with a protective earth contact. Do not negate
this protective action by the using an extension cord
without a protective conductor.
The following work must be carried out by a qualified
electrician. All local electrical codes must be strictly
observed. If the plug on the cable does not fit the power
outlet, or if the cable is to be attached to a terminal block,
cut the cable at the plug end and rewire it.
Cable color coding The color coding used in the cable depends on the cable
supplied.
Connecting a new plug If you are connecting a new plug, it should meet the local
safety requirements and include the following features:
Adequate load-carrying capacity (see table of
specifications).
Ground connection.
Cable clamp.
WARNING To avoid the possibility of injury or death, please note
that the Agilent E6000C does not have a floating earth.