User`s guide
Test I. Dynamic Range D Single-Mode/Multimode Module Performance Tests
E6000C Mini-OTDR User’s Guide, E0302
223
Figure 74 Dynamic Range Test Setup: Single-Mode
Figure 75 Dynamic Range Test Setup: Multimode
NOTE Instead of the 62.5m recirculating delay line, you can use 
a 62.5
µm multimode fiber of length > 4km. 
If you use such a multimode fiber, you do not require the 
coupler (within the dotted box in Figure 75).
2 Turn on the OTDR, and after the selftest has passed, 
recall the default settings.