D Single-Mode/Multimode Module Performance Tests Test III. Attenuation Deadzone
236 E6000C Mini-OTDR User’s Guide, E0302
♦ Or (Multimode module)
Set the pulsewidth to 10 ns.
[S
ETTINGS]<PULSEWIDTH><10 NS>. Close by OK.
Attenuator settings 8 Set up the attenuator.
–Set 
λ to the actual wavelength.
– Set the attenuation to 38 dB (see Note on page 235).
– Enable the attenuator output.
Start measurement 9 On the OTDR start the measurement.
Run/Stop
Position marker A 10 Wait until the backscatter noise is reduced (about 10 s), 
then position marker A close to the beginning of the 
first reflection after the front reflection.
U
P (A/B) until only A is highlighted. Use LEFT/RIGHT 
keys
Stop measurement 11 Stop the measurement.
Run/Stop
Set start 12 Set the start position close to the position of marker A 
and the measurement span to 2 km.
[S
ETTINGS]<RANGE><RANGE INPUT...>. Use Cursor keys to 
specify Start and Span. Confirm with OK.
NOTE The start position should be a little before the front edge 
of the reflection. For example, if the reflection is at 2.2 
km, use a start position of 2 km.
Restart measurement 13 Start the measurement.
Run/Stop