User`s guide

OTDR Mode 1 Getting Started
E6000C Mini-OTDR Users Guide, E0302
45
Relative to both markers The following parameters are measured between marker
A and marker B. The recorded values change when you
move either marker.
A-B: the distance between the markers
One of the following (selectable in the [A
NALYSIS]
menu):
2pt.L: 2-point loss between the markers. This is the
difference in power level between the marker points
2pt.Attn.: 2-point attenuation. This is the 2-point
loss per length unit.
LSA-Attn.: LSA Attenuation. This is the least
square approximation for the fiber loss per length
unit between the markers.
ORL: Optical Return Loss. This represents the
fraction of power reflected back to your Mini-OTDR.
Relative to current marker The following parameters show values at the current
marker. The recorded values change when you move or
change the current marker.
Ins.L. at A/B: the insertion loss of the event close
to the marker.
Refl. at A/B: the return loss (in dB) of the event
close to the marker.
Cum.L. to A/B: the cumulative loss between the
initial backscatter value interpolated to the start of the
fiber, and the marker point.
General parameters The following parameters are independent of the marker
position.
Range: the start position and the measurement span,
selectable from the [S
ETTINGS] menu. The available
Ranges are module-dependent.