Averaging Time Avg. Time: Averaging time of a measurement. The
measurement is stopped automatically when this time
has elapsed. Larger Averaging Times increase the
dynamic range by reducing the noise floor of the OTDR.
The specified dynamic range is reached after 3
minutes.
NOTE This parameter can also be configured to be Number of
Averages: a specified number of measurement
acquisitions. Number of Averages is a power of 2.
You specify the parameter used for Averaging in the OTDR
Settings page of the Instrument Configuration
pages (see “How to Set the OTDR Settings” on page 134).
Optimize Mode Optimize Mode: Resolution for short fibers,
Dynamic for long fibers, or Standard as a
compromise between Resolution and Dynamic.
Maximum Data Points DataPoints: the maximum number of data points. A
high value improves the resolution of the trace, but may
limit the number of traces that you can store in the
internal flash disk.
Front Connector Threshold FrontC. Thres: the Front Connector Threshold. This
is a threshold for reflectance of the Front Connector.
If reflectance is above this threshold, you receive a
warning message, saying Front Connector check
failed. If you see this message, you should clean your
Connector.
NOTE If you have chosen Reflection Height (see note below and
“How to Set the OTDR Settings” on page 134), the Front
Connector Threshold is not adjustable.
Refl. Thres: the Reflectance Threshold. Events with
a Reflectance above this threshold are displayed in the
Event Bar and Event Table.