The Settings screen 1 Getting Started
E6000C Mini-OTDR User’s Guide, E0302
53
Averaging Time • Avg. Time: Averaging time of a measurement. The 
measurement is stopped automatically when this time 
has elapsed. Larger Averaging Times increase the 
dynamic range by reducing the noise floor of the OTDR. 
The specified dynamic range is reached after 3 
minutes.
NOTE This parameter can also be configured to be Number of 
Averages: a specified number of measurement 
acquisitions. Number of Averages is a power of 2. 
You specify the parameter used for Averaging in the OTDR 
Settings page of the Instrument Configuration 
pages (see “How to Set the OTDR Settings” on page 134).
Optimize Mode • Optimize Mode: Resolution for short fibers, 
Dynamic for long fibers, or Standard as a 
compromise between Resolution and Dynamic.
Maximum Data Points • DataPoints: the maximum number of data points. A 
high value improves the resolution of the trace, but may 
limit the number of traces that you can store in the 
internal flash disk.
Front Connector Threshold • FrontC. Thres: the Front Connector Threshold. This 
is a threshold for reflectance of the Front Connector.
If reflectance is above this threshold, you receive a 
warning message, saying Front Connector check 
failed. If you see this message, you should clean your 
Connector.
NOTE If you have chosen Reflection Height (see note below and 
“How to Set the OTDR Settings” on page 134), the Front 
Connector Threshold is not adjustable.
• Refl. Thres: the Reflectance Threshold. Events with 
a Reflectance above this threshold are displayed in the 
Event Bar and Event Table.