User`s guide

The Settings screen 1 Getting Started
E6000C Mini-OTDR Users Guide, E0302
57
Total Link Loss Total Link Loss: Loss over whole fiber. This is
calculated as the loss between the Horizontal Offset
(see “How to Set the Horizontal Offset on page 96) and
the Fiber End.
An End Event must be present before this test can be
performed. See “How to Set the Fiber End” on page 98.
If the Loss between the Horizontal Offset and the Fiber
End is greater than this limit, this is reported in the
Pass/Fail test table.
The Total Link Loss limit can be anything up to 50dB.
Enter a value of 0 dB to deactivate this test. If the limit
is not active, you see a value of -.- dB in the Total
Link Loss edit field.
Attenuation Limit Attenuation: Attenuation Limit. If the LSA
attenuation between any 2 Events is greater than this
limit, the first Event is reported in the Pass/Fail test
table.
The Attenuation limit can be anything up to 5.000dB/
km. Enter a value of 0 dB/km to deactivate this test. If
the limit is not active, you see a value of -.-- dB/km
in the Attenuation edit field.
New Events New Events: Check for new events.
If you select New Events, the Pass/Fail test compares
the current trace with the most recent locked Event
Table. If the Pass/Fail test finds any Events which do