User`s guide
1 Getting Started The Settings screen
58 E6000C Mini-OTDR User’s Guide, E0302
not appear in the Event Table, they are reported in the 
Pass/Fail test table. 
If you select New Events, you should also set at least 
one other parameter in the Pass/Fail Param. 
window.
This feature is best used with a locked Event Table. See 
“How to Lock the Event Table” on page 91.
The check for New Events uses the current Scan Trace 
thresholds.
Link Length • Link Length: Distance to Fiber End. This is 
calculated as the difference between the Horizontal 
Offset (see “How to Set the Horizontal Offset” on 
page 96) and the Fiber End. 
An End Event must be present before this test can be 
performed. See “How to Set the Fiber End” on page 98.
If the Fiber End ± Horizontal Offset is more than 
Length Tolerance km (see below) from the Link Length, 
this is reported in the Pass/Fail test table.
In other words, if the Link Length is 100km, and the 
Length Tolerance is 2km, the recorded fiber length 
must be between 98km and 102km, otherwise you see 
an entry in the Pass/Fail test table.
The Link Length limit can be anything up to 500 km. 
Enter a value of 0 km to deactivate this test. If the limit 
is not active, you see a value of ---- km in the Link 
Length edit field.
Length Tolerance • Length Tolerance: Accepted margin of error used 
for checking the Link Length (see above). 
The Length Tolerance limit can be anything up to 50 
km. Enter a value of 0 dB to deactivate this test. If the 