How to Make an Automatic Measurement 3 Sample Sessions: Measuring a Trace
E6000C Mini-OTDR Users Guide, E0302
85
Automatic Settings 3 Cursor up to <AUTO> and press SELECT. Automatic
settings are now enabled.
You see the text Auto in the Range and PulseWidth
boxes, and the Mini-OTDR selects suitable settings for
your fiber.
4 Exit the S
ETTINGS menu by selecting OK.
Automatic Scanning 5 Select [V
IEW] from the popup panel. You see a list of
menu options.
6 If there is a tick next to <A
UTO SCAN>, Automatic
Scanning is already enabled. If Automatic Scanning is
not enabled. move down to <A
UTO SCAN> and press
S
ELECT or cursor right.
7 Leave the menu by cursor left, or selecting <C
LOSE VIEW
M
ENU>.
Start Measurement 8 Press the R
UN/STOP hardkey.
The light behind the R
UN/STOP hardkey goes on. After a
short initializing phase, the OTDR displays the first
result.
9 Press the R
UN/STOP hardkey, or wait for the end of the
measurement time, as indicated in the lower right
corner.
The light behind the R
UN/STOP hardkey goes off. No
more samples are being taken.
The OTDR now generates an Event Table and displays
the Event Table and Event Bar, if you have requested
them from the [V
IEW] menu.
NOTE If you have a color Mini-OTDR (E6000C option 003), you
can select whether or not the current display is color by
V
IEW - PREFERENCES options.
You can choose between B
LACK/WHITE, COLOR INDOOR (for
indoor use) and C
OLOR OUTDOOR (for outdoor use).