How to Run a Manual Measurement 3 Sample Sessions: Measuring a Trace
E6000C Mini-OTDR Users Guide, E0302
87
bottom left of the screen. Press SELECT to bring up the
next page.
Optimize Mode 3 Move to <O
PTIMIZE MODE> and press SELECT. You see
three options: <S
TANDARD>, <RESOLUTION> and
<D
YNAMIC>.
4 If you want to increase the dynamic range of the
measurement, move to <D
YNAMIC> and press SELECT.
5 Exit the Settings menu by selecting OK.
NOTE Parameters changed in the [S
ETTINGS] page only affect
subsequent traces. The current trace is unaltered.
The parameter values displayed on the Trace Screen
always refer to the current trace. Any parameter that has
been changed for subsequent traces is grayed.
NOTE If you alter the settings while a measurement is running,
press R
UN/STOP to start a new trace with the parameters
that you have just set.
How to Run the Measurement
Now that you have set the range correctly, the
measurement can be run:
1 Press the blue R
UN/STOP hardkey.
2 Wait for the trace to become free of noise. This takes
some seconds. Alternatively, wait until the
measurement time expires.
3 Press the R
UN/STOP hardkey.