User`s guide

How to Use the Pass/Fail test 3 Sample Sessions: Measuring a Trace
E6000C Mini-OTDR Users Guide, E0302
93
select Events for masking 5 Cursor to the Event(s) to be masked (or unmasked),
and press S
ELECT.
6 When all required Events are selected, select Ok.
NOTE To select all Events for masking, select Do All.
To deselect all events select Do All again.
Masked Events are not checked by subsequent Pass/Fail
tests. This means that these Events will not appear in the
Pass/Fail Test table.
Masked Events are indicated by a x between the entries
for No. and Type in the Event Table.
NOTE You can mask an individual event by selecting <M
ASK AN
E
VENT> from the [EVENTS] menu.
7 Select OK to exit the Measurement Settings screen.
How to run the Pass/Fail test
8 If you don’t already have a trace loaded, take a trace as
normal, either using the R
UN/STOP key, or by opening
an existing file using <O
PEN..> from the FILE menu.
Lock Event Table 9 If you are checking for New Events, lock the Event
Table.
See “How to Lock the Event Table” on page 91.
10 Select <P
ASS/FAIL TEST> from the [ANALYSIS] menu.
The Pass/Fail test checks the current trace against
limits set above. You see a message Pass/Fail Test
Active.., at the top of your Trace screen.
NOTE If you have not set any Pass/Fail Param. Limits, no Pass/
Fail test is performed.