User`s guide

3 Sample Sessions: Measuring a Trace How to Set the Horizontal Offset
96 E6000C Mini-OTDR Users Guide, E0302
The number of the Event at which the limit has been
exceeded (Relat. Evt. No.).
For Link Loss and Link Length (over the whole fiber),
this is the End Event.
Printout The Pass/Fail test table is also included in a printout. See
“How to Print the Measurement” on page 100.
NOTE If you want to perform a Pass/Fail test whenever a Scan
Trace is performed, select <A
UTO PASS/FAIL> from the
[V
IEW] menu.
NOTE If you want to view the Pass/Fail test table, or to stop
viewing it, select <P
ASS/FAIL TABLE> from the [VIEW]
menu.
You cannot see the Pass/Fail test table and the Event
Table simultaneously. This means that selecting the Pass/
Fail test table deselects the Event Table, and vice versa.
How to Set the Horizontal Offset
You use the Horizontal Offset to set all distances (for
example the marker position, or locations in the Event
Table) relative to this point. You do this as follows.
Position marker 1 Move your marker to the point where you want to set
the offset.
If you want to precisely position the marker, press the
D
OWN Cursor to view around the marker.
2 Select the menu option [V
IEW]<SET H-OFFSET to A>