User`s guide

How to Set the Fiber End 3 Sample Sessions: Measuring a Trace
E6000C Mini-OTDR Users Guide, E0302
99
Select End Threshold 3 Cursor to End Thres. and press SELECT. Follow “How
to change a numerical setting” on page 132 to select a
new threshold value.
If you select an end threshold of, for example, 3.0 dB,
an End will be set at the first event with an insertion
loss of 3 dB or more. If you select a threshold of 0 dB,
no End will be set.
Scan Trace 4 Select [A
NALYSIS]<SCAN TRACE> to run a scan trace.
The first Event which exceeds the specified End
Threshold is now set to type End, and subsequent
Events are ignored.
Or
1 Use the Cursors to move the current marker to an
Event.
Select [E
VENTS]<DECLARE END> from the popup panel.
An End is set at the event near the current marker.
NOTE If the current marker is not at an event, no End is set.
End Event The end Event is listed as type End in the Event table, and
marked on the Event Bar with a special symbol (see
Figure 32). All events to the right of the End Event are
removed