Product specifications

This series of linear-regulated
80-100 W DC power supplies is
designed to maximize the throughput
of DUTs through the manufacturing
test process. Both programming
and measurement are optimized for
speed. The active downprogrammer
can sink up to the full rated current
of the power supply, which quic kly
brings the power supply output to
zero volts. The 6630B series offers
many advanced programmable
features including stored states
and status reporting. Programming
is done using industry standard
SCPI commands via the GPIB or
RS-232. Test syst em i ntegration is
further simplified by using the
VXIPlug&Play drivers. The optional
relays simplify system design and
troubles hooting.
The optional front panel binding
posts make the 6630B series
convenient on the R&D lab bench.
The built-in microamp measurement
system helps the engineer to easily
and accurately monitor the output
voltage and current without a
complicated test setup.
Application notes
10 Practical Tips You Need to
Know About Your Power Products
5965-8239E
10 Hints for Using Your Power Supply
to Decrease Test Time
5968-6359E
High Performance DC Power Supplies speed and accuracy for test optimization
Sp ecifica ti on s
(at 0° to 55°C unless
otherwise specified)
6631B 6632B 6633B 6634B
More detailed specifications at
www.agilent.com/find/6630
40
Fast, low-noise outputs
Programmable active down-programmer sinks the full rated current
Dual-range, precision low current measurement
Optional isolation and polarity reversal relays
Built-in measurements and advanced programmable features
Protection features to ensure DUT safety
6631B - 6634B
Single-Output
80-100 W GPIB
Number of outputs 1 1 1 1
GPIB
Yes Yes Yes Yes
Output ratings
Voltage 0 to 8 V 0 to 20 V 0 to 50 V 0 to 100 V
Current 0 to 10 A 0 to 5 A 0 to 2 A 0 to 1 A
Programming accuracy at 25°C ±5°C
Voltage 5 mV 10 mV 20 mV 50 mV
+ Current 0.05% + 4 mA 2 mA 1 mA 0.5 mA
Ripple and noise (20 Hz to 20 MHz, with outputs
ungrounded or with either terminal grounded)
Voltage normal mode rms 0.3 mV 0.3 mV 0.5 mV 0.5 mV
peak-to-peak 3 mV 3 mV 3 mV 3 mV
Fast mode rms 1 mV 1 mV 1 mV 2 mV
peak-to-peak 10 mV 10 mV 15 mV 25 mV
Current rms 3 mA 2 mA 2 mA 2 mA
DC measurement accuracy via GPIB or front panel
meters with respect to actual output at 25°C ±5°C
Voltage 0.03% + 2 mV 3 mV 6 mV 12 mV
Low current range 20 mA to +20 mA 0.1% + 2.5 µA 2.5 µA 2.5 µA 2.5 µA
High current range +20 mA to + rated I 0.2% + 1 mA 0.5 mA 0.25 mA 0.25 mA
–20 mA to –rated I 0.2% + 1.6 mA 1.1 mA 0.85 mA 0.85 mA
Load regulation
Voltage 2 mV 2 mV 4 mV 5 mV
Current 2 mA 1 mA 1 mA 1 mA
Line regulation
Voltage 0.5 mV 0.5 mV 1 mV 1 mV
Current 1 mA 0.5 mA 0.25 mA 0.25 mA
Transient response time Less than 100 µs (50 µs in the fast mode) for the output voltage to recover to its previous
level (within 0.1% of the voltage rating of the supply or 20 mV) following any step change in load current of up to
50% of the output current rating of the supply.
Understanding Linear Power Supply
Operation (AN1554)
5989-2291EN