Operating instructions
 Language Dictionary
36
 Command Syntax
  *RCL <NRf>
 Parameters
 0 through 6 where:
 States 1-6 Volatile states previously stored by *SAV
 State 0 Nonvolatile state previously stored by *SAV 0
 Suffix
 None
 Related Commands
 *RST *SAV
 *RDT Resource Description Transfer Query
 Type
 Device  Specification
 Description
 This query returns the model number of a single electronic load or the model number of the
module installed in each channel of a multiple electronic load. For multiple electronic
loads, a semicolon (;) separates each module and the string terminated with a LF (line feed).
 Query Syntax
 *RDT?
 Returned Parameters
 < aard> String value as follows:
 single electronic load CHAN1:nnnnL; where “nnnnL” = model number
 multiple electronic load CHAN <c>:nnnnnL; where “c” = channel number and
  “nnnnnL” = number and suffix letter of the module in that
 channel.
 Related Commands
 *IDN? *OPT?
 *RST Reset Command
 Type
 Device  State
 Description
 This command sets the electronic load to its factory-defined state. (Refer to “Factory
Default Settings” in the Operating Manual of the electronic load model that you are
programming.) There are no parameters with this command; it sets all channels of the
multiple electronic load to the same state.
 *RST also does the following:
■  Forces an ABORt command before resetting any parameters.
■  After all parameters have been reset, executes an INP:PROT:CLE to clear the
electronic load’s protection circuits.
 Note
 *RST does not affect any Status Enable registers or Transition Filters.
 Command Syntax
 *RST
 Parameters
 None
 Related Commands
 *RCL *SAV