 Query Syntax
 *SRE?
 Returned Parameters
 <NR1> Value: 0 to 255
 Suffix
 None
 Related Commands
 *PSC
 *STB? Read Status Byte Query
 Type
 Device  Status
 Description
 This query reads the Status Byte register. Note that the MSS (Master Summary Status) bit
and not the RQS bit is returned in Bit 6. This bit indicates whether or not the electronic
load has at least one reason for requesting service. *STB? does not clear the Status Byte
register, which is cleared only when subsequent action has cleared all its set bits. Refer to
Chapter 5 - Status Reporting for more information about this register.
  Status Byte Register
 Bit  Position
 7 6 5 4  3  2 1 0
 Condition
 OPER
 MSS
 ESB  MAV  QUES
 CSUM  
1
0 
1
0
 Bit  Weight
 128  64  32  16  8  4  2  1
1
Always zero.
 Parameters
 None
 Returned Parameters
 <NR1> Value: 0 to 255
 Suffix
 None
 Related Commands
 None
 *TRG Immediate Trigger Command
 Type
 Device  Trigger
 Description
 This command which is essentially the same as the Group Execute Trigger (<GET>),
generates a trigger to the electronic load only if TRIG:SOUR is set to BUS (refer to the
TRIGger Subsystem Root commands).
 Command Syntax
 *TRG
 Parameters
 None
 Related Commands
 <GET> TRIG:SOUR