Operating instructions
Language Dictionary
41
ABORt Channel- Independent Termination Command
 Description
 This command applies only to trigger functions. It cancels all pending [:LEVel]:TRIG
operations (such as CURR:TRIG) in all operating modes and on all channels. As a result,
subsequent triggers have no effect on the input level. This command resets the WTG bit of
the Operation Condition register (refer to Chapter 5 - Status Reporting) and has the same
effect on status as the receipt of a trigger. ABORt has no affect on the Transient
Subsystem.
 Command Syntax
 ABORt
 Parameters
 None
 Examples
  ABOR  Aborts all pending trigger level operations in the Current,
Resistance, or Voltage Subsystems
 CURR 10; CURR: TRIG 20;: ABOR;: TRIG  Cancels the programmed trigger level so 
that the triggered input remains at 10 amps
 Query
 None
 Related Commands
 CURR[:LEV]:TRIG VOLT[:LEV]:TRIG STAT:OPER:COND?
CHANnel Channel-Independent Command/Query
 Note
 This command is for the multiple electronic load but is supported in single electronic loads
for compatibility.
 Description
 CHANnel selects the multiple electronic load channel to which all subsequent channel-
specific commands will be directed. If the specified channel number does not exist or is
outside the MIN/MAX range, an error code is generated (See Table 4-2 at the end of this
chapter). When used with the single electronic load the only valid parameter is 1.
 Note
 This command and INSTrument are the only channel commands for electronic loads. No
other methods of channel selection are supported.
 Command Syntax
 CHANnel[:LOAD] <NRf+>
 Parameters
 See Table 4-1 and the Operating Manual of the electronic load model.
 Examples
 CHANNEL:LOAD 5  Select Channel 5
 CHAN 5
 CHAN MAX Select highest existing channel in a multiple electronic load.
 Query Syntax
 CHAN? CHAN? MAX CHAN? MIN