 Command Syntax
 INPut:SHORt[:STATe]  <NRf+>
 Parameters
   Value  Range  Units  *RST  Default  
   OFF  or 0
 ON  or 1
 None
 None
 0
 Examples
 INP:SHOR ON  Set shorted input condition
 INP:SHOR  1
 CHAN 2;INP:SHOR 1;:CHAN 1;INP:SHOR 0
 On multiple electronic load, short Channel 1 and unshort Channel 2
 Query Syntax
 INP:SHOR?
 Returned parameters
 <NR1> Value; 0 for unshorted, 1 for shorted
 Related Commands
 INP[:STAT]
INP[:STATe] Channel-Specific Input Command/Query
 Description
 This implied keyword turns the electronic load input ON and OFF. When INPut is OFF,
the electronic load will draw minimal current at its input. INP[STAT]e:OFF overrides
INPput:SHORt:ON. The presently programmed slew rate setting is not used when turning
the input on or off; the current and voltage change at their maximum rates .
 Command Syntax
 INPut [:STATe] <NRf+>
 Parameters
   Value  Range  Units  *RST  Default  
   OFF  or 0
 ON  or 1
 None
 None
 0
 Examples
 INP:STAT OFF  Turn electronic load input off
 INP  0
 INP:STAT OFF;SHOR ON;:INP ON Turn input off, short the input, and turn input on.
 Query Syntax
 INP?
 Returned Parameters
 <NR1> Value; 0 for OFF, 1 for ON
 Related Commands
 CURR:PROT INP:SHOR
MEASure Channel-Specific Measurement Query
 Description
 This function consists of queries that return the current, voltage, and power at the input of
the electronic load.
 Query Syntax
  Query
 MEASure:CURRent[:DC]?
 MEASure:POWer[:DC]?
 MEASure:VOLTage[:DC]?
  Value Returned
 electronic load input current
 Computed electronic load input power
 electronic load input voltage