Operating instructions
Language Dictionary
53
 Parameters
 Enter the desired mode as a string variable in either the full or abbreviated format. There
are no units. The *RST Default is CURRent.
 Examples
 MODE :CURR  Set electronic load input to CC mode
 CHAN 1;: MODE:CURR; :CHAN 2; :MODE:RES Set Channel 1 input to CC mode 
and Channel 2 input to CR mode
 Query Syntax
 MODE?
 Returned Parameters
 <aard> String value: CURR, RES, or VOLT
 Alternate Syntax
 FUNCtion can be used as an alias for MODE.
 Syntax Diagram
 Mode Command Syntax Diagram
PORT0 Channel-Specific Output Command/Query
 Description
 This command controls the general-purpose digital port on the rear of each electronic load
or each channel of the multiple electronic load. The state of the port is with respect to the
common side of the electronic load input. Setting a port to the OFF or 0 state connects it
to the common; setting a port to the ON or 1 state disconnects it from common and sets it
to a TTL logical high.
 The command is the same for both single and multiple electronic loads. To control a port
in a multiple electronic load you must first select the channel and then use PORT0 on that
channel.
 Command Syntax
 PORT0[:STATe] <NRf+>
 Note
 This command does not have a four-character short form. You must always include the
fifth character (zero).
 Parameters
   Value  Range  Units  *RST  Default  
   OFF  or 0
 ON  or 1
 None
 None
 0