Operating instructions
 Language Dictionary
72
TRlGger Subsystem Channel-Independent Trigger Programming Function
 Description
 In a multiple electronic load, triggering can be initiated from the controller, the a-c line,
the external Trigger jack, or the mainframe timer. single electronic loads may be triggered
from their external Trigger jack, or from the controller. Any trigger signal generated by a
single electronic load or multiple electronic load also exists at the Trigger output jack and
may be used as a trigger source for any other electronic load.
 Note
 For a detailed explanation of the effects of triggers, see Operation Overview in the
electronic load Operating Manual.
 Keywords
 Command
 TRIGger[:IMMediate]
 Function
 Trigger-true input from the controller.
   TRIGger:SOURce EXTernal
 Selects trigger source at external Trigger jack.
   TRIGger:SOURce BUS
 Selects GPIB *TRG or <GET> as trigger source.
   TRIGger:SOURce HOLD
 Disables all trigger inputs except the TRlG:IMM command.
1
TRIGger:SOURce LINE
 Selects the internal a-c line frequency detector as the trigger source.
1
TRIGger:SOURce TIMer
 Selects internal trigger oscillator as the trigger source.
1
 These commands apply only to multiple electronic loads
 Related Commands
 *TRG <GET>
 Related Subsystems
 TRANsient Subsystem