Technical data
120 6000 Series Oscilloscopes Service Guide
5 Replacing 6000A Assemblies
8 If the e- field shield requires replacement, use a 
screwdriver to pry the shield off the studs.
Figure 29 Removing the backlight inverter shield
9 To reinstall the backlight inverter board:
a Align the shield over the studs and push it all the way 
down on the studs.
b Reconnect the LCD cables looping them around each 
other and through the cable guides as shown in 
Figure 28.
c Reconnect the supply cable.
d Align the holes in the inverter board with the posts on 
the display mount and push down until the latches 
snap over the board (refer to Figure 28). 
e Route the cables through the cable guides as shown.
Studs
6000_series_service_guide.book Page 120 Friday, March 23, 2007 2:43 PM