User Calibration
Perform user-calibration:
• Each year or after 2000 hours of operation.
• If the ambient temperature is >10° C from the calibration 
temperature.
• If you want to maximize the measurement accuracy.
The amount of use, environmental conditions, and 
experience with other instruments help determine if you 
need shorter User Cal intervals.
User Cal performs an internal self- alignment routine to 
optimize the signal path in the oscilloscope. The routine uses 
internally generated signals to optimize circuits that affect 
channel sensitivity, offset, and trigger parameters. Disconnect 
all inputs and allow the oscilloscope to warm up before 
performing this procedure.
Performing User Cal will invalidate your Certificate of 
Calibration. If NIST (National Institute of Standards and 
Technology) traceability is required perform the procedures 
in  Chapter 2 in this book using traceable sources.
To perform User Cal
1 Set the rear-panel CALIBRATION switch to 
UNPROTECTED.
2 Connect short (12 inch maximum) equal length cables to 
each analog channel’s BNC connector on the front of the 
oscilloscope. You will need two equal-length cables for a 
2- channel oscilloscope or four equal-length cables for a 
4- channel oscilloscope.