Technical data
84 6000 Series Oscilloscopes Service Guide
4 Troubleshooting
Equipment required for troubleshooting
The equipment listed in this table is required to troubleshoot the 
oscilloscope.
Table 13 Equipment Required to Troubleshoot the Oscilloscope
To check out the oscilloscope
1 Disconnect any external cables from the front panel. 
2 Disconnect the power cord, then remove the cabinet 
following the instructions on page 104 for Agilent 6000A 
Series models or page 145 for Agilent 6000L Series 
models.
Equipment Critical Specifications Recommended Model/Part 
Digital multimeter Accuracy ±0.05,% 1 mV resolution Agilent 34401A 
Oscilloscope Capable of measuring ≥ 500 MHz signal. 
1 M
Ω input impedance.
Agilent 
DSO6102A, MSO6102A, 
DSO6102A, or MSO6104A
CAUTION
USE AN EXTERNAL FAN TO AVOID OVERHEATING COMPONENTS !
When you remove the oscilloscope cover and main shield, use an external fan to 
provide continuous air flow over the heat sinks. Air flow over the heat sinks is 
reduced when the cover and main shield are removed, which leads to higher than 
normal operating temperatures. Have the fan blow air across the system board 
where the heat sinks are located. Otherwise, damage to the components can 
occur.
If the cover of a 6000A Series oscilloscope is removed but the main shield remains 
installed and the bottom holes are not blocked, the instrument will cool properly. 
WARNING
HAZARDOUS VOLTAGES EXIST — HIGH VOLTAGE IS PRESENT ON DSO6000L 
SERIES POWER SUPPLY HEAT SINKS !
The power supply heat sinks of the 6000L Series oscilloscopes are at a high 
potential. This presents an electric shock hazard. Protect yourself from electric 
shock by keeping this area covered or by not coming in contact with the heat sinks 
when the power cord is attached to the oscilloscope! 
6000_series_service_guide.book Page 84 Friday, March 23, 2007 2:43 PM










