Technical data
Troubleshooting 4
6000 Series Oscilloscopes Service Guide 95
4 Perform a continuity check on the “DC cable” (item W1 
on  page 167). If the cable is shorted or open, replace 
it. If not, continue to the next step.
5 Reconnect the “DC cable” to the power supply, but not 
to the system board.
6 Connect the power cord and check the power supply 
voltage between pins 1 and 6 of the system board side 
of the “DC cable”. The voltage should be 15 V ±10%. 
• If the voltage is correct, then test the system board 
(see  page 96). 
• If the voltage is not correct, then replace the power 
supply 
(see  page 150). 
WARNING
HAZARDOUS VOLTAGES EXIST — HIGH VOLTAGE IS PRESENT ON DSO6000L 
SERIES POWER SUPPLY HEAT SINKS !
The power supply heat sinks of the 6000L Series oscilloscopes are at a high 
potential. This presents an electric shock hazard. Protect yourself from electric 
shock by keeping this area covered or by not coming in contact with the heat sinks 
when the power cord is attached to the oscilloscope! 
6000_series_service_guide.book Page 95 Friday, March 23, 2007 2:43 PM