User`s guide

Table Of Contents
Absolute Measurement Examples
6
Agilent E5505A User’s Guide 189
The appropriate measurement definition parameters for this example
have been pre-stored in this file. Table 28 on page 200 lists the
parameter data that has been entered for the Microwave Source
measurement example.)
5
Using Figure 133 on page 190 as a guide, navigate to the Sources tab.
a
Enter the carrier (center) frequency of your DUT (5 MHz to 1.6 GHz).
Enter the same frequency for the detector input frequency.
b
Enter the VCO Tuning Constant (see Table 26 on page 190). Use the
following equation to calculate the appropriate VCO Tuning Constant to
enter for the measurement.
VCO Tuning Constant = T x Carrier Frequency, where T= 5E-9
For example, to calculate the Tuning Constant value to enter for EFC tuning
when the center frequency is 18 GHz:
(5 E – 9) X (18 E + 9) = 90
c
Enter the Tune Range of VCO (see Table 26 on page 190).
d
Enter the Center Voltage of VCO (see Table 26).
e
Enter the Input Resistance of VCO (see Table 26).
NOTE
Note that the source parameters in Ta b l e 28 on page 200 may not be appropriate for the
reference source you are using. To change these values, refer to Table 26 on page 190,
then continue with step 5 below. Otherwise, go to the section “Beginning the
measurement" on page 193.