Residual Measurement Examples
8
Agilent E5505A User’s Guide 241
3
Press the Continue button when ready to calibrate the measurement.
4
Adjust the phase difference at the phase detector as prompted by the phase
noise software. See Figure 175.
.
5
The system will measure the positive and negative peak voltage of the phase
detector using an internal voltmeter. The quadrature meter’s digital display
can be used to find the peak. The phase may be adjusted either by varying
the frequency of the source or by adjusting a variable phase shifter or line
stretcher.
6
The system software will then prompt you to set the phase noise software’s
meter to quadrature by adjusting the phase shifter until the meter indicates
0 volts, then press Continue. See Figure 176 on page 242.
Figure 175 Adjust phase difference at phase detector