AM Noise Measurement Fundamentals
11
Agilent E5505A User’s Guide 293
3
Locate the drive level on the AM sensitivity graph (Figure 224), and enter
the data.
4
Measure the noise data and interpret the results. The measured data will be
plotted as single-sideband AM noise in dBc/Hz.
Figure 223 AM noise calibration setup
NOTE
The quadrature meter should be at zero volts due to the blocking capacitor at the AM
detector’s output.
Figure 224 AM detector sensitivity graph
e5505a_user_AM_cal.ai
rev2 10/24/03