Agilent Technologies InfiniiVision 7000B Series Oscilloscopes Data Sheet Engineered for the best signal visibility
If you haven’t purchased an Agilent scope lately, why should you consider one now? Oscilloscopes are visual tools and larger, high-resolution displays have become increasingly important as general purpose scopes need more space to display digital and serial signals in addition to traditional scope channels.
What gives the InfiniiVision 7000B Series the best signal visibility? 1. Biggest display Oscilloscopes are visual tools and larger, high-resolution screens make the product better. Bigger displays have become increasingly important as general purpose scopes need more space to display digital waveforms and serial decode traces in addition to traditional analog waveforms. The 12.1-inch display size helps you easily view up to 20 channels simultaneously with serial protocol. 2.
Your design has analog, digital and serial signals ... shouldn’t your scope? Analog: Up to 1 GHz bandwidth and up to 4 GSa/s sample rate The InfiniiVision 7000B Series scope channels provide faster identification of your most elusive problems – Revolutionary high-resolution display. Engineered with an XGA display and 256 levels of intensity grading, see a precise representation of the analog characteristics of the signals you’re testing.
Other useful features Dedicated Front Panel Key Navigate Through the Captured Waveform Automatically marks each frame in the selected search criteria. Use the dedicated front panel navigation keys to automatically jump to the next or previous marked frame or play through the entire waveform with control to stop, rewind or adjust the speed of waveform scrolling. Simply press the dedicated front panel Search Key and enter a variety of specific search criteria High resolution mode.
Why does a fast update rate matter? While bandwidth, sample rate and memory depth are key criteria for deciding which scope to purchase, an equally important characteristic is update rate. What is update rate? Update rate is how many waveforms acquisitions per seconds your scope can acquire, process, and display. Oscilloscope “dead-time” is the time it takes for a scope to process and then display an acquired waveform before re-arming it’s triggering for the next acquisition.
How update rate affects signal visibility Capturing random and infrequent events on an oscilloscope is all about statistical probabilities. The key to improving the probability of capturing a signal anomaly is to minimize dead-time and take more pictures of the signal in a given timeframe. Here is an example with Tek and Agilent scopes both connected to a target with a glitch that occurs 25 times per second.
Software applications Mask testing (N5455A or Option LMT) Agilent’s mask test option (Option LMT or N5455A) for InfiniiVision Series oscilloscopes provides a fast and easy way to test your signals to specified standards, and uncover unexpected signal anomalies, such as glitches. Mask testing on other oscilloscopes is based on softwareintensive processing technology, which tends to be slow. Agilent’s InfiniiVision scopes can perform up to 100,000 real-time waveform pass/fail tests per second.
Software applications CAN/LIN triggering and decode (N5424A or Option AMS on new scope purchases) Trigger on and decode serially transmitted data based on CAN and LIN protocols. This application not only provides triggering on complex serial signals, but it also provides unique hardware-accelerated capabilities. Hardware-accelerated triggering and decode means the scope stays responsive and fast.
Software applications and other accessories Power application (U1881A) Need to make power measurements with your scope? Agilent’s power application provides a full suite of power measurements that run on a PC connected to an InfiniiVision 7000B Series oscilloscope. Make more accurate power supply efficiency measurements by using an U1880A deskew fixture to deskew your voltage and current probes. Use your scope to quickly make and analyze power measurements. For more information: www.agilent.
Probes and accessories Agilent offers a complete family of innovative passive and active probes for the InfiniiVision 7000B Series scopes helps you get your job done easily and accurately. Choosing the correct probe for your application will help to ensure you accurately acquire signals. Below is a general guide on how to choose the type of probe. For the most up-to-date information about Agilent’s accessories, please visit our Web site at www.agilent.com/find/scope_probes.
Connectivity The 7000 B Series scopes come with the most comprehensive connectivity tools in their class. LXI class C LAN eXtensions for Instrumentation (LXI) is a standards-based architecture for test systems. By specifying the interaction of system components, LXI enables fast and efficient test system creation and reconfiguration.
Agilent InfiniiVision Portfolio Agilent’s InfiniiVision lineup includes 2000X, 3000X, 6000, and 7000B Series oscilloscopes. These share a number of advanced hardware and software technology blocks. Use the following selection guide to determine which scope best matches your specific needs.
Agilent InfiniiVision 7000B Series oscilloscopes: Engineered for the best signal visibility 12.1” large display makes it easier to view analog, digital and serial signals. High-resolution color display with XGA resolution and 256 levels of intensity reveals subtle details that most scopes won’t show you. Free IntuiLink data capture PC software makes transferring waveform data or a screen image to a PC fast and easy. Built-in Web viewer via LAN allows for remote measurements and viewing.
Search and Navigate front panel controls make it easy to find a view specific signal activity. Quickly pan and zoom for analysis with MegaZoom III’s instant response and optimum resolution. Autoscale lets you quickly display any analog or digital active signals, automatically setting the vertical, horizontal and trigger controls for the best display, while optimizing memory.
Performance characteristics Acquisition: scope channels Sample rate MSO/DSO701xB: 2 GSa/s each channel MSO/DSO703xB: 2 GSa/sec each channel MSO/DSO705xB, 710xB: 4 GSa/sec half channel*, 2 GSa/sec each channel Equivalent-time sample rate: 400 GSa/s (when real-time mode is turned off) Maximum memory depth Standard 2 channels/4 channels 8 Mpts/4 Mpts Vertical resolution 8 bits Peak detection MSO/DSO701xB: 500-ps peak detect MSO/DSO703xB: 500-ps peak detect MSO/DSO705xB/710xB: 250-ps peak
Performance characteristics (continued) Vertical system: scope channels Scope channels MSO/DSO7xx2B: Ch 1 and 2 simultaneous acquisition MSO/DSO7xx4B: Ch 1, 2, 3 and 4 simultaneous acquisition Bandwidth (–3 dB)* MSO/DSO701xB: DC to 100 MHz MSO/DSO703xB: DC to 350 MHz MSO/DSO705xB: DC to 500 MHz MSO/DSO710xB: DC to 1 GHz AC coupled MSO/DSO701xB: 3.5 Hz to 100 MHz MSO/DSO703xB: 3.5 Hz to 350 MHz MSO/DSO705xB: 3.5 Hz to 500 MHz MSO/DSO710xB: 3.5 Hz to 1 GHz Calculated rise time (=0.
Performance characteristics (continued) Vertical system: scope channels (continued) ESD tolerance ±2 kV Noise, RMS, input shorted MSO/DSO701xB: 0.50% FS or 300 µV, whichever is greater MSO/DSO703xB: 0.50% FS or 300 µV, whichever is greater MSO/DSO705xB: 0.50% FS or 360 µV, whichever is greater MSO/DSO710xB: 0.65% FS or 360 µV, whichever is greater DC vertical gain accuracy*1 ±2.0% full scale DC vertical offset accuracy ≤200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value; >200 mV/div: ±0.
Performance characteristics (continued) Horizontal Range MSO/DSO701xB: 2 nsec/div to 50 sec/div MSO/DSO703xB: 2 nsec/div to 50 sec/div MSO/DSO705xB: 1 nsec/div to 50 sec/div MSO/DSO710xB: 500 psec/div to 50 sec/div Resolution 2.
Performance characteristics (continued) Trigger system (continued) Selections Edge Pattern Edge, pulse width, pattern, TV, duration, sequence, CAN, LIN, USB, I2C, SPI, Nth edge burst, RS-232 with Option 232 Trigger on a rising, falling, alternating or either edge of any source Trigger at the beginning of a pattern of high, low, and don’t care levels and/or a rising or falling edge established across any of the analog and digital channels, but only after a pattern has stabilized for a minim
Performance characteristics (continued) Trigger system (continued) Nth edge burst Trigger on the Nth edge of a burst that occurs after an idle time that you specify. Max edge count: 65,536. Autoscale Finds and displays all active scope and logic (for 7000B Series MSO) channels, sets edge trigger mode on highest-numbered channel, sets vertical sensitivity on scope channels and thresholds on logic channels, time base to display ~1.8 periods. Requires minimum voltage >10 mVpp, 0.
Performance characteristics (continued) Display system Display 12.1-inch (255 mm x 184 mm) diagonal color TFT LCD Throughput of scope channels Up to 100,000 waveforms/sec in real-time mode Resolution XGA – 768 vertical by 1024 horizontal points (screen area); 640 vertical by 1000 horizontal points (waveform area) 256 levels of intensity scale Controls Waveform intensity on front panel.
Performance characteristics (continued) FFT Points Up to 128 Kpts in precision mode Source of FFT 1, 2, 1+2, 1-2, 1x2, MSO/DSO7xx4A: 3, 4, 3+4, 3-4, 3x4; where 1, 2, 3, 4 represent the analog channel inputs 1, 2, 3, and 4 Window Rectangular, flattop, Hanning, Blackman Harris Noise floor –50 to –90 dB depending on averaging Amplitude Display in dBV, dBm at 50 Ω Frequency resolution 0.
Performance characteristics (continued) General characteristics (continued) Trigger out When Triggers is selected (delay ~17 ns) 0 to 5 V into high impedance 0 to 2.
Ordering information Model Bandwidth Sample rate Memory depth Scope channels DSO7012B 100 MHz DSO7014B MSO7012B Digital channels 2 GSa/s 8 Mpts 2 100 MHz 2 GSa/s 8 Mpts 4 100 MHz 2 GSa/s 8 Mpts 2 16 MSO7014B 100 MHz 2 GSa/s 8 Mpts 4 16 DSO7032B 350 MHz 2 GSa/s 8 Mpts 2 DSO7034B 350 MHz 2 GSa/s 8 Mpts 4 MSO7032B 350 MHz 2 GSa/s 8 Mpts 2 16 MSO7034B 350 MHz 2 GSa/s 8 Mpts 4 16 DSO7052B 500 MHz 4 GSa/s 8 Mpts 2 DSO7054B 500 MHz 4 GSa/s 8 Mpts 4 MSO705
Ordering information (continued) Options Product Description DSO to MSO upgrade* N2741A for DSO701xB N2735A for DSO703xB N2736A for DSO705xB N2737A for DSO710xB SEC Secure Environment Mode - Provides compliance with National Industrial Security Program Operating Manual (NISPOM) Chapter 8 requirements (factory-installed option only for new purchase) A6J ANSI Z540 compliant calibration *Includes a 54620-68701 logic cable kit, a label and an upgrade license to
Agilent Technologies Oscilloscopes Multiple form factors from 20 MHz to > 90 GHz | Industry leading specs | Powerful applications 27
www.agilent.com www.agilent.com/find/7000B myAgilent myAgilent www.agilent.com/find/myagilent A personalized view into the information most relevant to you. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Agilent is a founding member of the LXI consortium. Agilent Channel Partners www.agilent.