Specifications

RF-IV instruments require calibration every time the instrument is powered on or every time the
frequency setting is changed. This is because ambient temperature, humidity, frequency setting, etc.
have a much greater influence on measurement accuracy than in low frequency impedance mea-
surements. Calibration is performed using open, short, and load reference terminations (a low loss
capacitor termination is also used as necessary) as described in Section 4.7.1. The calibration plane
is at the test port or the tip of test port extension where the calibration reference terminations are
connected (see Figure 4-2.)
Note: The calibration of the RF I-V instruments that should be performed prior to measurements
eliminates impedance measurement errors under the desired measurement conditions. The
RF I-V instruments also require periodic calibration at the recommended intervals for
maintaining their overall operating performance within specifications.
Figure 4-2. Calibration plane of RF-IV instruments
RF impedance/ material analyzer
(a) Open/short/load (+ LLC) calibration at test port
Calibration plane
Open
Short
Load
(b) Open/short/load (+ LLC) calibration at the tip of a port
extension cable
Calibration plane
Open
Short
Load
(c) Open/ short/ load calibration at DUT
contact terminals of a test fixture
Calibration plane
Test fixture
Open
(No device)
Short
Load
LLC: Calibration using low loss
capacitor termination
4-2