Technical data

1 Installation
10 Agilent Multimode for 6210/6220 TOF Installation and Maintenance Guide
8 In the MS TOF tab, click Parameters and select Negative ion polarity. Click
Save As to save the negative tune parameters as d:\Program Files\Agilent\
TOF Software\Tune\NegativeESI.prm.
9 Right-click on the TOF icon in the Status pane and click Vent.
10 After the TOF system has completed the vent process, turn off the front
switch and the main breaker.
11 Prepare a static-free work surface to store the electronics assemblies that
are removed during the upgrade procedure.
Step 2. Remove the TOF E-Tub Electronics
Remove the following assemblies from the TOF Electronics Tub in this
order:
1 Top Foam piece from the Electronics Tub.
2 RF Power Amplifier and front and back foam pieces.
3 Mid Foam piece from the Electronics Tub
4 Analyzer 3 PCA (G1946-65250)
5 SmartCard 4+ PCA with PDB cover
6 Power Distribution Board (G1946-65002)