Technical data

1 Installation
28 Agilent Multimode for 6210/6220 TOF Installation and Maintenance Guide
\
Figure 25 Capillary
8 After cleaning the capillary and cap, place the capillary cap back on
capillary. Install the new spray shield with field shaping electrodes.
Figure 26 End Cap Assembly, Multimode (p/n G1978-60060)
9 Screw the multimode spray shield into the holder for the spray shield. See
Figure 27.
Note that the field shaping electrodes are in the 9 o'clock and 6 o'clock
positions.
10 Loosen the end plate screws on each side to adjust the field shaping
electrodes position.
Capillary