Technical data

7
Agilent G1978A Multimode for 6210/6220 TOF
Set-Up Guide
Agilent Technologies
1
Installation
Installation 9
Step 1. Prepare to install 9
Step 2. Remove the TOF E-Tub Electronics 10
Step 3. Change the firmware chips on the Analyzer and Power
Distribution Boards 11
Step 4. Install the three 10MΩ cables 14
Step 5. Install the HV control PCA and cables 16
Step 6. Attach the multimode source to the instrument 21
Step 7. Install the nebulizer, corona needle, and gas connections 24
Changing Sources 26
To convert between the ESI, APCI or APPI and the multimode ion
source 26
This chapter contains instructions to install the G1978A Multimode Source.
You need to make changes to the instrument as part of the installation.
The multimode source is supported with the following software:
TOF Software A.02.00 or A.02.01 with G1978-10004 patch
MassHunter Workstation Acquisition for TOF A.02.02
MassHunter Workstation Acquisition for TOF B.01.03 or higher
To upgrade from A.02.02 to B.01.03 or higher, you must completely remove the
Analyst software. To make sure Analyst is completley removed from your
system, reimage the computer.
Refer to the TOF Installation Guide for software installation instructions.