Specifications
Calibration 101
CAL:CURR:MON (Series 668xA/669xA only) 
This command sets the power supply to the current monitor (I
MON
) calibration point that is then entered with 
CAL:CURR[:DATA]. The output current must be calibrated after CAL:CURR:MON is performed. 
 Command Syntax   CALibrate:CURRent:MONitor 
 Parameters   <NRf+> 
Examples  CAL: CURR: MON CALibrate: CURRent: MONitor 
Query Syntax 
(None) 
Related Commands  CAL:CURR[:DATA] CAL:STAT 
CAL:PASS 
This command enters a new calibration password. The command is active only when the power supply is already in the 
calibration mode. Unless it is changed subsequently to shipment, the password is the power supply's four-digit model 
number. If the password is set to 0, password protection is removed and 
CAL:STAT ON is unrestricted. A new password is 
automatically stored in nonvolatile memory and does not have to be stored with the 
CAL:SAVE command. 
Command Syntax 
 CALibrate:PASScode <NRf> 
Parameters 
<NRf> 
Examples 
CAL:PASS 6671 CAL:PASS 09.1993 
Query Syntax 
(None) 
Related Commands 
CAL:STAT 
CAL:SAVE 
This command saves any new calibration constants (after a current or voltage calibration procedure has been completed) in 
nonvolatile memory. 
 Command Syntax: 
 CALibrate:SAVE 
 Parameters 
(None) 
Examples 
CAL: SAVE 
Query Syntax 
(None) 
Related Commands 
CAL:CURR CAL:VOLT CAL:STAT 
CAL:STAT 
This command enables and disables the calibration mode. The calibration mode must be enabled before the power supply 
will accept any other calibration commands. The first parameter specifies the enabled or disabled state. The second 
parameter is the password. It is required if the calibration mode is being enabled and the existing password is not 0. If the 
second parameter is not entered or is incorrect, an error is generated and the calibration mode remains disabled. The query 
statement returns only the state, not the password. 
Whenever the calibration mode is changed from enabled to disabled, any new calibration constants are lost unless they have 
been stored with CAL:SAVE. 
Command Syntax: 
CALibrate:STATe <bool> [,<NRf>] 
Parameters 
{0 | OFF} | {1 | ON} [,<NRf>] 
*RST Value 
OFF 
Examples 
CAL:STAT 1,6671 CAL:STAT OFF 
Query Syntax 
CALibrate:STATe? 
Returned Parameters 
{0 | 1 } 
Related Commands 
CAL:PASS CAL:SAVE 