Specifications
General Information 36
Table 1-5a. Performance Specifications for Series 669xA
1
Parameter  Agilent Model Number 
 6690A 6691A 6692A   
Output Ratings 
Voltage: 
0 - 15 V  0 - 30 V  0- 60 V 
Current:* 
0 - 440 A  0 - 220 A  0 – 110 A 
 *Derated linearly 1%/°C from 40 ° C to 55 °C 
Programming Accuracy (@ 25 ± 5 °C) 
Voltage: 0.04% + 
15 mV  30 mV  60 mV 
Current: 0. l % + 
230 mA  125 mA 
65 mA 
Ripple & Noise (from 20 Hz to 20 MHz with outputs ungrounded, or with either output terminal grounded)  
Constant Voltage: rms 
2.5 mV  2.5 mV  2.5 mV 
Constant Voltage: p-p 
15 mV  25 mV  25 mV 
 Readback Accuracy (from front panel or over GPIB with respect to actual output @ 25 ± 5 °C)  
Voltage: 0.05% + 
22.5 mV  45 mV  90 mV 
±
±±
±Current 0.1% + 
300 mA  165 mA  80 mA 
Load Regulation (change in output voltage or current for any load change within ratings) 
Voltage 0.002% + 
650 µV 
1.1 mV  2.2 mV     
Current: 0.005% + 
40 mA  17 mA  9 mA     
Line Regulation (change in output voltage or current for any line change within ratings) 
Voltage: 0.002% + 
650 µV 650 µV 650 µV 
Current: 0.005% + 
40.5 mA  17 mA  9 mA     
Transient Response Time (for the output voltage to recover to within 150 mV following any step change from 100% to 
50% or 50% to 100% of the rated output current): < 900 µs 
AC Line Input * (selectable - see Appendix C) 
Range 1 (180-235 Vac) 
 Nominal phase-to-phase voltage: 
 Input frequency: 
200, 208 Vac (3-phase) 
50/60 Hz 
Range 2 (360-440 Vac) 
 Nominal phase-to-phase voltage: 
 Input frequency: 
400, 416 Vac (3-phase) 
50/60 Hz 
* Power source can be DELTA or WYE. 
Notes: 
1
For Supplemental Characteristics, see Table 1-5b. 