Specifications
General Information  39
Table 1-5b. Supplemental Characteristics for Series 669xA (continued)
1
Parameter All Models 
Command Processing Time (Average time for output voltage to change after receipt of digital data when the supply is 
connected directly to the GPIB Bus): 20 ms 
Monotonicity: 
Output is monotonic over entire rated voltage, current, and 
temperature range. 
Auto-Parallel Configuration: 
Up to 3 identical models 
Nonvolatile Storage 
 State storage & recall locations: 
 Prestored turn-on state: 
 Maximum memory write cycles: 
4 
Location 0 
40,000, typical 
Digital Port Characteristics 
(see Table 1-6) 
GPIB Interface Capabilities 
(see Table 1-6) 
Serial Link Capabilities 
(see Table 1-6) 
Recommended Calibration Interval: 
1 year 
Safety Compliance 
 Complies with: 
 Designed to comply with: 
CSA 22.2 No.231, 
IEC 1010 (carries CE mark) 
UL 1244 
RFI Suppression (complies with): 
CISPR-ll, Group 1, Class B 
Dimensions 
Width: 
 Height 
 including removable feet 
 excluding removable feet 
 Depth (without output safety cover): 
425.5 mm (16.75 in) 
234.2 mm (9.25 in) 
221.5 mm (8.75 in) 
674.7 mm (25.56 in) 
Weight 
 Net: 
 Shipping: 
51.3 kg (113 lb) 
63.6 kg (140 lb) 
Output Characteristic Curve: 
Notes: 
l
For Performance Specifications, see Table l-4a. 
Vmax
Vset
CV operating
line
0
1
2
Vout
Iout
CC operating line
Agilent
Model
Vout
Iout
6690A
6691A
6692A
15V
30V
60V
440A
220A
110A
Maximun Rated Output
Output Characteristic Curve:
Iset
Imax