User`s guide
 Language Dictionary - 8
103
 SENSe:FUNCtion
 This command configures the measurement sensor to measure either voltage or current when an acquire
trigger is used. The query returns the function setting, either VOLT or CURR.
Command Syntax
 SENSe:FUNCtion <function>
 Parameters
 "VOLTage" | "CURRent"
 Examples
 SENS:FUNC "VOLT"
 Query Syntax
 SENSe:FUNCtion?
 Returned Parameters
 <SRD>
SENSe:PROTection:STATe
 This command enables or disables the open sense lead detection.
Command Syntax
 SENSe:PROTection:STATe <bool>
 Parameters
 0 | OFF | 1 | ON
 *RST Value
 1
 Examples
 SENS:PROT:STAT ON
 Query Syntax
 SENSe:PROTect:STATe?
 Returned Parameters
 <NR3>
SENSe:SWEep:OFFSet:POINts
 This command defines the offset in a data sweep when an acquire trigger is used. Negative values
represent data samples taken prior to the trigger. Positive values represent the delay after the trigger
occurs but before the samples are acquired.
Command Syntax
 SENSe:SWEep:OFFSet:POINts <NRf+>
 Parameters
 -4095 through 2,000,000,000
 *RST Value
 0
 Examples
 SENS:SWE:OFFS:POIN -2047
 Query Syntax
 SENSe:SWEep:OFFSet:POINts?
 Returned Parameters
 <NR3>
 Related Commands
 SENS:SWE:TINT SENS:SWE:POIN MEAS:ARR
SENSe:SWEep:POINts
 This command defines the number of points in a data sweep.
Command Syntax
 SENSe:SWEep:POINts<NRf+>
 Parameters
 1 through 4096
 *RST Value
 2048
 Examples
 SENS:SWE:POIN 1024
 Query Syntax
 SENSe:SWEep:POINts?
 Returned Parameters
 <NR3>
 Related Commands
 SENS:SWE:TINT SENS:SWE:OFFS MEAS:ARR