User`s guide
 8 – Language Dictionary
106
 OUTPut:PON:STATe
 This command selects the power-on state of the dc source. This information is saved in non-volatile
memory. The following states can be selected:
 RST
 Sets the power-on state to *RST. Refer to the *RST command as described in this
chapter for more information.
 RCL0
 Sets the power-on state to *RCL 0. Refer to the *RCL command as described in
this chapter for more information.
Command Syntax
 OUTPut:PON:STATe <state>
 Parameters
 RST | RCL0
 Examples
 OUTP:PON:STAT RST
 Query Syntax
 OUTPut:PON:STATe?
 Returned Parameters
 <CRD>
 Related Commands
 *RST *RCL
OUTPut:PROTection:CLEar
 This command clears the latch that disables the output when an overvoltage, overcurrent,
overtemperature, or remote inhibit status condition is detected. All conditions that generate the fault must
be removed before the latch can be cleared. The output is then restored to the state it was in before the
fault condition occurred.
Command Syntax
 OUTPut:PROTection:CLEar
 Parameters
 None
 Examples
 OUTP:PROT:CLE
 Related Commands
 OUTP:PROT:DEL *RCL *SAV
OUTPut:PROTection:DELay
 This command sets the time between the programming of an output change that produces a constant
current condition (CC) and the recording of that condition by the Operation Status Condition register.
The delay prevents the momentary changes in status that can occur during reprogramming from being
registered as events by the status subsystem. Since the constant current condition is used to trigger
overcurrent protection (OCP), this command also delays OCP. Overvoltage protection is not affected by
this command.
Command Syntax
 OUTPut:PROTection:DELay <NRf+>
 Parameters
 0 to 2,147,483.647
 Unit
 seconds
 *RST Value
 0.08 (Normal)
 Examples
 OUTP:PROT:DEL 75E-1
 Query Syntax
 OUTPut:PROTection:DELay?
 Returned Parameters
 <NR3>
 Related Commands
 OUTP:PROT:CLE *RCL *SAV