User`s guide
 8 – Language Dictionary
110
  [SOURce:]VOLTage:PROTection
 This command sets the overvoltage protection (OVP) level of the dc source. If the output voltage exceeds
the OVP level, then the dc source output is disabled and the Questionable Condition status register OV
bit is set (see chapter 7 about programming the status registers). An overvoltage condition can be cleared
with the OUTP:PROT:CLE command after the condition that caused the OVP trip is removed. The OVP
always trips with zero delay and is unaffected by the OUTP:PROT:DEL command. The OVP function
can be disabled using the VOLT:PROT:STAT command.
Command Syntax
 [SOURce:]VOLTage:PROTection[:LEVel]<NRf+>
 Parameters
 see Table 8-3
 Default Suffix
 V (volts)
 *RST Value
 MAX
 Examples
 VOLT:PROT 21.5 VOLT:PROT:LEV MAX
 Query Syntax
 [SOURce:]VOLTage:PROTection[:LEVel]?
 Returned Parameters
 <NR3>
 Related Commands
 OUTP:PROT:CLE VOLT:PROT:STAT
  [SOURce:]VOLTage:PROTection:STATe
 This command enables or disables the overvoltage protection (OVP) function.
CAUTION: Disabling the overvoltage protection function may cause excessive output voltages, such
as can occur if remote sense leads are disconnected, to damage the equipment under test.
Command Syntax
 [SOURce:]VOLTage:PROTection:STATe <bool>
 Parameters
 0 | 1 | OFF | ON
 *RST Value
 OFF
 Examples
 VOLT:PROT:STAT 0 VOLTAGE:PROTECTION:STATE OFF
 VOLT:PROT:STAT 1 VOLTAGE:PROTECTION:STATE ON
 Query Syntax
 [SOURce:]VOLTage:PROTection:STATe?
 Returned Parameters
 <NR1>0 or 1
 Related Commands
 OUTP:PROT:CLE *RST VOLT:PROT
  [SOURce:]VOLTage:TRIGger
 This command sets the pending triggered voltage level of the dc source. The pending triggered level is a
stored voltage value that is transferred to the output terminals when a trigger occurs. In order for a trigger
to occur, the trigger subsystem must be initiated (see the INITiate command in the trigger subsystem).
Command Syntax
 [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <NRf+>
 Parameters
 see Table 8-3
 Default Suffix
 V (volts)
 *RST Value
 0
 Examples
 VOLT:TRIG 20 VOLTAGE:LEVEL:TRIGGERED 20
 Query Syntax
 [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
 Returned Parameters
 <NR3>
 Related Commands
 VOLT *RST