User`s guide
 Language Dictionary - 8
93
 CALibrate:DATA
 This command enters a calibration value that you obtain by reading an external meter. You must first
select a calibration level (with CALibrate:LEVel) for the value being entered.
Command Syntax
 CALibrate:DATA<NRf>
 Parameters
 <external reading>
 Unit
 A (amperes)
 Examples
 CAL:DATA 3222.3 MA CAL:DATA 5.000
 Related Commands
 CAL:STAT CAL:LEV
 CALibrate:LEVel
 This command selects the next point in the calibration sequence.
 P1:  the first calibration point
 P2:  the second calibration point
Command Syntax
 CALibrate:LEVel <point>
 Parameters
 P1 | P2
 Examples
 CAL:LEV P2
 CALibrate:PASSword
 This command lets you change the calibration password. A new password is automatically stored in
nonvolatile memory and does not have to be stored with CALibrate:SAVE.
 If the password is set to 0, password protection is removed and the ability to enter the calibration mode is
unrestricted.
Command Syntax
 CALibrate:PASScode<NRf>
 Parameters
 <model number> (default)
 Examples
 CAL:PASS 6812 CAL:PASS 6.1994
 Related Commands
 CAL:SAV
 CALibrate:SAVE
 This command saves any new calibration constants after a calibration procedure has been completed in
nonvolatile memory. If CALibrate:STATe OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored..
Command Syntax
 CALibrate:SAVE
 Parameters
 None
 Examples
 CAL:SAVE
 Related Commands
 CAL:PASS CAL:STAT